Contributions
Abstract: PB2352
Type: Publication Only
Background
The cobas m 511 integrated hematology analyzer is a novel slide-based system that performs a CBC, WBC differential, reticulocyte count, and nucleated RBC count using digital analysis of a microscope slide.
Aims
To assess the accuracy of low platelet (PLT) counts on the cobas m 511 compared with the Sysmex® XN-10 Automated Hematology Analyzer and BD FACSCanto™ Flow Cytometer in a single-center study.
Methods
Residual whole blood samples (n=115) in the PLT range 1-505 x103/µL (based on flow cytometery using CD61, CD41a, and CD42b antibodies) were randomly processed on each system. Sysmex samples were set to automatically reflex to PLT-F function, which is marketed as more reliable for thrombocytic or problematic samples than the standard impedance PLT-I function. Three analyses were performed: cobas m 511 and Sysmex Analyzer versus flow cytometer (reference), and cobas m 511 versus Sysmex Analyzer (reference). Data were compared using three thresholds and results included if a valid PLT result was obtained on both systems.
Results
All low PLT samples had valid results with the cobas m 511 and flow cytometry. Two (1.7%) samples were flagged with the Sysmex Analyzer (PLT-F) indicating unreliable results. The cobas m 511 demonstrated excellent accuracy compared with flow cytometry, with ≥93.6% sensitivity, specificity, and agreement, and good accuracy compared with Sysmex Analyzer (Table 1). Attempts were made to also study results for the Sysmex Analyzer PLT-I function; however, the vast majority of samples (100% for PLT counts ≤10, 80% for PLT 11–20, and 28% for PLT 21–50 x103/µL) had flags indicating unreliable values using this function which prevented any further comparison. 
  | PLT threshold (103/µL)  | Sensitivity (%)  | Specificity (%)  | Agreement (%)  | 
cobas m 511 vs.   | 10  | 95.5  | 96.8  | 96.5  | 
20  | 93.6  | 94.1  | 93.9  | |
50  | 98.6  | 97.6  | 98.3  | |
Sysmex   Analyzer (PLT-F)  | 10  | 100.0  | 96.7  | 97.3  | 
20  | 100.0  | 92.5  | 95.6  | |
50  | 100.0  | 97.6  | 99.1  | |
cobas m 511 vs.  
  | 10  | 88.0  | 97.7  | 95.6  | 
20  | 94.1  | 100.0  | 97.3  | |
50  | 100.0  | 100.0  | 100.0  | 
Conclusion
These data demonstrate the robustness and accuracy of platelet counts reported by the cobas m 511 system. Many samples were flagged using the Sysmex PLT-I function that, in routine use, would have reflexed to the PLT-F function. All samples on the cobas m 511 system had valid PLT results.
Session topic: 33.  Platelets disorders
Keyword(s): Platelet
Abstract: PB2352
Type: Publication Only
Background
The cobas m 511 integrated hematology analyzer is a novel slide-based system that performs a CBC, WBC differential, reticulocyte count, and nucleated RBC count using digital analysis of a microscope slide.
Aims
To assess the accuracy of low platelet (PLT) counts on the cobas m 511 compared with the Sysmex® XN-10 Automated Hematology Analyzer and BD FACSCanto™ Flow Cytometer in a single-center study.
Methods
Residual whole blood samples (n=115) in the PLT range 1-505 x103/µL (based on flow cytometery using CD61, CD41a, and CD42b antibodies) were randomly processed on each system. Sysmex samples were set to automatically reflex to PLT-F function, which is marketed as more reliable for thrombocytic or problematic samples than the standard impedance PLT-I function. Three analyses were performed: cobas m 511 and Sysmex Analyzer versus flow cytometer (reference), and cobas m 511 versus Sysmex Analyzer (reference). Data were compared using three thresholds and results included if a valid PLT result was obtained on both systems.
Results
All low PLT samples had valid results with the cobas m 511 and flow cytometry. Two (1.7%) samples were flagged with the Sysmex Analyzer (PLT-F) indicating unreliable results. The cobas m 511 demonstrated excellent accuracy compared with flow cytometry, with ≥93.6% sensitivity, specificity, and agreement, and good accuracy compared with Sysmex Analyzer (Table 1). Attempts were made to also study results for the Sysmex Analyzer PLT-I function; however, the vast majority of samples (100% for PLT counts ≤10, 80% for PLT 11–20, and 28% for PLT 21–50 x103/µL) had flags indicating unreliable values using this function which prevented any further comparison. 
  | PLT threshold (103/µL)  | Sensitivity (%)  | Specificity (%)  | Agreement (%)  | 
cobas m 511 vs.   | 10  | 95.5  | 96.8  | 96.5  | 
20  | 93.6  | 94.1  | 93.9  | |
50  | 98.6  | 97.6  | 98.3  | |
Sysmex   Analyzer (PLT-F)  | 10  | 100.0  | 96.7  | 97.3  | 
20  | 100.0  | 92.5  | 95.6  | |
50  | 100.0  | 97.6  | 99.1  | |
cobas m 511 vs.  
  | 10  | 88.0  | 97.7  | 95.6  | 
20  | 94.1  | 100.0  | 97.3  | |
50  | 100.0  | 100.0  | 100.0  | 
Conclusion
These data demonstrate the robustness and accuracy of platelet counts reported by the cobas m 511 system. Many samples were flagged using the Sysmex PLT-I function that, in routine use, would have reflexed to the PLT-F function. All samples on the cobas m 511 system had valid PLT results.
Session topic: 33.  Platelets disorders
Keyword(s): Platelet
                            
        